Technical data
104 N9320A Programmer’s Guide
5 Command Reference
bandwidth, and the occupied bandwidth results is displayed
in the data window. If the power of the adjacent channel is
included in the calculation for the 100% power, the OBW
result will be too high. The OBW Span should be set narrow
enough to encompass the channel of interest and exclude
any unwanted adjacent channels.
*RST: 3 MHz
Example: OBW:FREQ:SPAN 10 MHz
Key access:
> Occupied BW > OBW Span
Turn on/off Max Hold
[:SENSe]:OBWidth:MAXHold OFF|ON|0|1
[:SENSe]:OBWidth:MAXHold?
Enables you to turn maximum hold trace feature On or Off for
the measurement. Maximum hold displays and holds the
maximum responses of a signal.
*RST: Off
Example: OBW:MAXH ON
Key access: > Occupied BW > Max Hold
Set the Occ BW% Pwr
[:SENSe]:OBWidth:PERCent <percent>
[:SENSe]:OBWidth:PERCent?
Enables you to change the percentage of signal power used
when determining the occupied bandwidth.
*RST: 99.0%
Range: 10% - 99.99%
Example: OBW:PERC 50
Key access: > Occupied BW > Occ BW% Pwr
Meas
BW/
Avg
BW/
Avg