Technical data

102 N9320A Programmer’s Guide
5 Command Reference
[:SENSe]:CORRection Subsection
External Amplifier Correction
[:SENSe]:CORRection:OFFSet[:MAGNitude]
<rel_ampl>
[:SENSe]:CORRection:OFFSet[:MAGNitude]?
A single value of amplitude correction can be applied to the
displayed trace data to compensate for signal losses or gains
that are due to other devices in the measurement setup,
rather than the unit under test.
*RST: 0 dB
Range: –81.9 to 81.9
Default Unit: dB
Key Access: > More > Ext Amp Gain
[:SENSe]:OBWidth Subsection
Set the Average Number On/Off
[:SENSe]:OBWidth:AVERage:COUNt <integer>
[:SENSe]:OBWidth:AVERage:COUNt?
[:SENSe]:OBWidth:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:OBWidth:AVERage[:STATe]?
To specify the number of measurement averages used when
calculating the measurement result set Avg Number to ON. The
average will be displayed at the end of each sweep. Setting
Avg Number to OFF disables measurement averaging.
*RST: 10/Off
Range: 1 to 1000
Example: OBW:AVER COUN 20
OBW:AVER OFF
Key access:
> Occupied BW > Avg Number On Off
Amplitude
Meas