Technical data
Command Reference 5
N9320A Programmer’s Guide 95
• EXP (Exponential Averaging mode) — When you set Avg
Mode to Exp, each successive data acquisition after the
average count is reached is exponentially weighted and
combined with the existing average. Exponential averaging
weights new data more than old data, which facilitates
tracking of slow- changing signals. The average will be
displayed at the end of each sweep.
• Repeat — When you set Avg Mode to Repeat, after reaching
the average count, all previous result data is cleared and
the average count is set back to 1.
*RST: EXPonential
Example: ACP:AVER:TCON EXP
Key access:
> ACP > Avg Mode
Set the Channel Integration BW
[:SENSe]:ACPower:BANDwidth|BWIDth:INTegration
<freq>
[:SENSe]:ACPower:BANDwidth|BWIDth:INTegration?
Specify the range of integration used in calculating the
power in the main channel.
*RST: 2.0 MHz
Default Unit: Hz
Range: 300 Hz - 20 MHz
Example: ACP:BWID:INT 5E6
Key access:
> ACP > Chan Integ BW
Set ACP Reference Auto
[:SENSe]:ACPower:CARRier:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:ACPower:CARRier:AUTO[:STATe]?
Enables you to set the adjacent channel power reference to
automatic or manual. When set to automatic, the carrier
power result reflects the measured value in the carrier.
Meas
Meas