User`s guide

72 N9320A Spectrum Analyzer
4 Key Reference
Amplitude
Press hardkey to activate the reference level function
and access the amplitude softkey menu.
The softkey menu selections are described as follows:
Ref Level Softkey sets the reference level. This function is activated
when is pressed. The reference level is the amplitude
power or voltage represented by the top line of the graticule
on the display. Setting the reference level changes the
absolute amplitude level (in the selected amplitude units) of
the top graticule line.
Key access:
Attenuation
Auto Man
Softkey toggles the input attenuation between Auto and Man.
The attenuation level can be set in as little as 1 dB
increments when Man is selected. The analyzer input
attenuator is normally coupled to the reference level control,
it reduces the power level of the input signal at the input
mixer. Selecting Auto re- couples the attenuator. Change the
attenuation by using the arrow keys, the knob, or the
numeric keypad.
Key access:
Scale/Div Softkey sets the logarithmic units per vertical graticule
division on the display. The Scale/Div function is available
only when the Scale Type key is set to Log.
Key access:
Scale Type
Log Lin
Softkey scales the vertical graticule divisions in logarithmic
units when Log is selected. The valid logarithmic units are 1,
2, 5 and 10 dB per division. When Lin is underlined, the
vertical divisions are linearly scaled with default amplitude
units of volts. The top of the graticule represents the
reference- level value; the bottom of the graticule represents
zero volts. (Each division is one- tenth of the reference level.)
Key access:
Amplitude
Amplitude
Amplitude
Amplitude
Amplitude
Amplitude