User`s guide
Functions and Measurements 3
N9320A Spectrum Analyzer 29
Note that the frequency and amplitude of the marker appear
both in the active function block, and in the upper- right
corner of the screen.
Figure 1 10 MHz Internal Reference Signal
2 Use the knob, the arrow keys, or the softkeys in the Peak
Search menu to move the marker. The marker information
will be displayed in the upper- right corner of the screen.
Changing Reference Level
1 Press and note that reference level (Ref Level) is now
the active function.
2 Press
> Mkr-> Ref Lvl.
Marker
10.000000 MHz
0.43 dBm
Active function block
Marker Annotation
Amplitude
Marker
NOTE
Changing the reference level changes the amplitude value of the top
graticule line.