User`s guide

28 N9320A Spectrum Analyzer
3 Functions and Measurements
Viewing a Signal
Refer to the procedures below to view a signal.
1 Press > Pow on/Preset > Preset Type > Factory to enable
the factory- defined preset state.
2 Press > Preset to restore the analyzer to its
factory- defined state.
3 Connect the 10 MHz REF OUT on the rear panel to the
front- panel RF IN.
Setting the Reference Level and Center Frequency
1 Press > 10 > dBm to set 10 dBm reference level.
2 Press > 30 > MHz to set 30 MHz to center frequency.
Setting Frequency Span
Press > 50 > MHz to set 50 MHz frequency span.
Reading Frequency and Amplitude
1 Press to place a marker (labeled 1) on the 10 MHz
peak.
Preset/
System
Preset/
System
Amplitude
Frequency
SPAN
NOTE
Changing the reference level changes the amplitude value of the top
graticule line. Changing the center frequency changes the horizontal
placement of the signal on the display. Increasing the span will
increase the frequency range that appears horizontally across the
display.
Peak
Search