User`s guide

Functions and Measurements 3
N9320A Spectrum Analyzer 27
Presetting the Spectrum Analyzer
Preset function provides a known instrument status for
making measurements. There are two types of presets,
factory and user:
Factory Preset When this preset type is selected, it restores the analyzer to
its factory- defined state. A set of known instrument
parameter settings defined by the factory. Refer to “Factory
Preset State” on page 111 for details.
User Preset Restores the analyzer to a user- defined state. A set of user
defined instrument parameter settings saved for assisting the
user in quickly returning to known a instrument
measurement setup.
Press > Pwr on/Preset > Preset Type to select the preset
type.
When Preset Type is set to Factory, pressing > Preset
triggers a factory preset condition. The instrument will
immediately return to the factory default instrument
parameter setting.
When Preset Type is set to User, pressing > Preset
displays both User Preset and Factory Preset softkeys. The user
may then select the preset desired from the softkey menu
selections.
Setting up a User Preset
To quickly return to instrument settings that are user
defined, perform the following steps to save the instrument
state as the user- defined preset:
1 Set the instrument parameters to the values and settings
necessary for the user preset state. This would include the
frequency, span, amplitude, BW, and measurement type and
any other setup details desired.
2 Press > Pwr on/Preset > Save User Preset, to save the
current instrument settings as the ‘user preset’ state. The
user preset will not affect the default factory preset settings.
User preset settings can be changed and saved at any time.
Preset/
System
Preset/
System
Preset/
System
Preset/
System