User`s guide
22 N9320A Spectrum Analyzer
2 Getting Started
Some helpful Tips
The following contains information to help in using and
maintaining the instrument for optimum operation, including
alignment, external reference, firmware update and option
activation.
Running Internal Alignments
To meet the instrument performance specifications, the
analyzer must periodically be manually aligned.
1 Connect a BNC cable with a correct type N- BNC adapter
between the CAL OUT and RF IN front panel connectors.
2 After instrument warm- up, press > Alignment > Align >
All.
Refer to “Alignment” on page 112 for details.
Enable the Options
Option license key information is required to enable product
options. Contact your nearest Agilent Office for purchasing a
license. Refer to the procedures below to activate the options
you have purchased. (option TG3 and option PA3)
1 Press > More > More > Licensing > Option
2 Enter the option number to be enabled. Press [Enter] to
confirm your input.
3 Press > More > More > Licensing > License key
4 Enter the license key information. Press [Enter] to confirm
your input and terminate the license key input. The option
will be enabled immediately.
NOTE
When an alignment is being run, there will be an audible clicking
sound as the attenuator settings are changed. This sound is not an
indication of a problem.