User`s guide

148 N9320A Spectrum Analyzer
6 Instrument Messages
767 Fatal LDS error
Attempt to initialize the instrument state has failed. Cycle
instrument power. If this fails to correct the problem,
contact your nearest Agilent Technologies service center.
768 Unable to load user state
Attempt to load a state failed, because the state was
saved with a
measurement personality that is not currently
loaded. Load the appropriate personality and try again.
770 Incorrect filename
You sent a remote command to save a screen file but did not
specify a valid extension.
780 No Peak Found
No signal peak was found that meets the criteria under
> Search Criteria.
781 Video Trigger cannot be active
You sent a remote command to do one of the following:
Turn on video trigger while the Average Detector or a
Marker Function is active.
Turn on the Average Detector or a Marker Function while
Video Trigger is active.
782 Video Trigger cannot be used with Mkr Func
Due to the Average Detector, the video Trigger cannot be
used with Maker function.
783 Allowable span for current CF exceeded
784 Allowable CF for current span exceeded
805 Doing Alignment
Peak
Search