User`s guide
One-Button Measurements 5
N9320A Spectrum Analyzer 137
Res BW
Auto Man
Specifies the resolution bandwidth used in measuring the
offset pair. When set to Auto, the Res BW from the default
tables for “Offset/Limits” on page 134 are used. When set to
Man, the allowed range is 1 kHz to 1MHz.
Key access:
> More > Spectrum Emission Mask > Offset/Limits
Abs Start To enter an absolute level limit at Start Freq for the currently
selected offset ranging from –200.00 to +50.00 dBm with 4
significant digits.
Key access:
> More > Spectrum Emission Mask > Offset/Limits
> More
Abs Stop
Couple Man
To enter an absolute level limit at Stop Freq for the currently
selected offset ranging from –200.00 to +50.00 dBm with 4
significant digits, and to toggle this function between Couple
and Man. If set to Couple, this is coupled to Abs Start to make
a flat limit line. If set to Man, Abs Start and Abs Stop the user
can enter different values to make a sloped limit line.
Key access:
> More > Spectrum Emission Mask > Offset/Limits
> More
Rel Start Enters a relative level limit at Start Freq ranging from –150.00
to +50.00 dBc with 4 significant digits.
Key access:
> More > Spectrum Emission Mask > Offset/Limits
> More
Rel Stop
Couple Man
Enters a relative level limit at Stop Freq ranging from –150.00
to +50.00 dBc with 4 significant digits, and to toggle this
function between Couple and Man. If set to Couple, this is
coupled to Rel Start to make a flat limit line. If set to Man, Rel
Start and Rel Stop the user can enter different values to make
a sloped limit line.
Key access:
> More > Spectrum Emission Mask > Offset/Limits
> More
Fail Mask Displays the menu to select one of the following logic keys
for fail conditions between the measurement results and the
test limits: Absolute, Relative, Abs AND Rel, and Abs OR Rel.
Key access:
> More> Spectrum Emission Mask > Offset/Limits
> More > More
Meas
Meas
Meas
Meas
Meas
Meas