User`s guide

One-Button Measurements 5
N9320A Spectrum Analyzer 131
When
Avg Mode is Repeat, after reaching the average count, all
previous data is cleared and the average count is back to 1.
Key access:
> More> Intermod (TOI)
TOI Span Specifies the frequency span in which intermodulation
products are measured. If you modify the value of span in
the base instrument Span menu, the value in the Intermod
(TOI) menu will be updated to reflect the new value, and the
measurement will restart if it is running.
Key access:
> More> Intermod (TOI)
Max Mixer Lvl
Auto Man
Sets the relationship between the highest signal that can be
displayed (the reference level) and the input attenuation. The
relationship applies whenever the Attenuation is set to Auto.
The relationship is that the attenuation is given by reference
level minus the max mixer level. For example, as the
reference level changes, the input attenuator changes to
ensure that a signal at the reference level does not exceed
the Max Mixer Lvl setting.
When the TOI measurement is off, the previous maximum
Max Mixer Lvl is restored.
Pressing Auto sets the maximum mixer level to –30 dBm.
Key access:
> More> Intermod (TOI)
Meas
Meas
Meas