User`s guide

Key Reference 4
N9320A Spectrum Analyzer 117
Sweep/Trig
Accesses the Sweep/Trigger softkeys for selection of the
sweep mode and trigger mode of the analyzer. The softkey
menu selections are defined as follows:
Sweep Time
Auto Man
Selects the length of time the analyzer takes to tune across
the displayed frequency span (or, in zero span, the time the
analyzer takes to sweep the full screen). Reducing the sweep
time increases the sweep rate. Change the sweep time by
using the arrow keys, the knob, or the numeric keypad.
In non- zero spans, when the sweep time is auto- coupled, the
analyzer selects the optimum (shortest) sweep time for the
current settings. This selection process is influenced by
several factors:
The max. tuning rate of the spectrum analyzer
The selected resolution and video bandwidth filters
The max. sample rate of the analog- to- digital converter
The number of trace points
The amplitude scale (logarithmic or linear)
The detector mode
The user may select a sweep time less than the auto- coupled
value; however this may generate measurement errors. The
minimum sweep time is 50 ms.
In zero span, the minimum sweep time is determined by the
maximum sample rate of the ADC and the number of sweep
points. The minimum sweep time is 2 microseconds. The
maximum sweep time is 4000 seconds.
Key access:
Sweep
Single
Pressing Sweep Single puts the analyzer in single- sweep
mode. Press Sweep Single to initiate a sweep when in
single- sweep mode.
Key access:
Sweep Cont Selecting continuous sweep causes repeated sweeps to be
triggered. Pressing Preset (When Preset Type is set to Factory)
sets the sweep mode to continuous sweep.
Sweep/
Tri g
Sweep/
Tri g