User`s guide

Key Reference 4
N9320A Spectrum Analyzer 103
Amptd Step
Auto Man
To set the step size of the power level range of the
tracking generator. The default setting is one vertical
scale division when in a logarithmic amplitude scale.
Key access:
> Tracking Generator
Power
Sweep
Sweep the output offset of the tracking generator.
The analyzer continues to sweep the linear change
between start and stop frequency when power sweep
is on. The input range is 0 dB to 10 dB.
Key access:
> Tracking Generator>More
Normalize Accesses the following softkeys.
Store Ref
(1->3)
Copies trace 1 into trace 3. Softkey menu
Store Ref (1->3) must be pressed before
pressing Normalize (On).
Key access:
> Tracking Generator >
More > Normalize
Normalize
On Off
Activates the normalize function. On
each sweep, the normalized trace (Trace
3) is subtracted from Trace 1 and the
result is added to the normalized
reference level.
The trace data is normalized with
respect to the normalized reference level,
even if the value of the normalized
reference level is changed. This function
remains in effect on all subsequent
sweeps until it is turned off.
The normalize function is most useful for
applying correction data to a trace while
making a stimulus response measurement
MODE
MODE
MODE
NOTE
Trace 1 should be in clear write mode prior to setting normalize to on.