Specifications

Calibrate Commands
CALibrate:STATe 0|OFF|1|ON [,<password>]
CALibrate:STATe?
Enables or disables calibration mode. Calibration mode must be enabled for the instrument to accept any calibration
commands. The first parameter specifies the state. The second optional parameter is the password.
Parameter Typical Return
0|OFF|1|ON , default OFF 0or1
<password>a numeric value up to 15 digits (none)
Disable calibration mode and unsecure calibration: CAL:STAT OFF,0
Enable calibration: CAL:STAT ON
<password> is optional to secure the instrument, but must be correct if provided.
CALibrate:VOLTage[:LOCal][:LEVel] [<value>]
Calibrates the local voltage programming and measurement. The optional value parameter selects the range to cal-
ibrate. This is for compatibility with the N6700 series modular power system.
Parameter Typical Return
Optional - the maximum value of the
range.
(none)
Calibrates the voltage of the 80 V range: CAL:VOLT 80
CALibrate:VOLTage:REMote[:LEVel] [<value>]
Calibrates the remote voltage programming and measurement. The optional value parameter selects the range to cal-
ibrate. This is for compatibility with the N6700 series modular power system.
Parameter Typical Return
Optional - the maximum value of the
range.
(none)
Calibrates the voltage of the 80 V range: CAL:VOLT:REM 80