Specifications
Programming the Output
Programming the Output
When the power supply is first turned on, it may take about 30 seconds or so to initialize the unit
before it is ready for use.
Set the Output Voltage
Set the Output Current
Enable the Output
Set the Output Voltage
You can set the output voltage when the output is either disabled (OFF) or enabled (ON). When enabled, the output
voltage will rise to the programmed setting. This assumes that a minimum output current has been programmed, that
the output has not reached its current limit or power limit boundary.
Front Panel SCPI Command
Turn the Voltage knob while observing the
voltage setting; or
Press the Voltage key, enter a value.
Then press Select.
To set the output voltage to 90 volts:
VOLT 90
When the unit is operating in constant voltage mode, CV status is displayed on the front panel.
Set the Output Current
You can set the output current when the output is either disabled (OFF) or enabled (ON). When enabled, the output cur-
rent will be limited at its programmed setting.
Front Panel SCPI Command
Turn the Current knob while observing
the current setting; or
Press the Current key, enter a Value.
Then press Select.
To set the output current to 100 amps:
CURR 100
When the unit is operating in constant current mode, CC status is displayed on the front panel.
Enable the Output
HAZARDOUS VOLTAGES All models generate voltages in excess of 60 VDC, with some
models rated at up to 1,500 VDC! Ensure that all instrument connections, load wiring,
and load connections are either insulated or covered so that no accidental contact
with lethal output voltages can occur.
Front Panel SCPI Command
Press the On/Off key. To turn the output on or off:
OUTP ON|OFF
Agilent N8900 Series Operating and Service Guide 55