Specifications
AC Mains Connections
Power Connector
Remove the strain relief to access the connector plug. Connect the AC mains cable to the connector plug as shown in
the following illustrations. Insert the cable wires into the connector plug. On 400 VAC connectors, insert a flat blade
screwdriver to release the internal wire clamp. On 208 VAC connectors, turn the screw to open or close the clamp.
Tighten all wires securely. The safety ground wire must be green. Other colors are for illustration purposes.
400 VAC Connector
Maximum wire size: 8 AWG or 10 mm
2
Maximum insulation diameter: 6.8 mm
Wire stripping length: 13 - 15 mm
Optional ferrule suggestion:
For 8 AWG: WAGO 216-289
For 10 AWG: WAGO 216-288
For 12 AWG: WAGO 216-287
For 14 AWG: WAGO 216-286
208 VAC Connector
Maximum wire size: 6 AWG or 16 mm
2
Maximum insulation diameter: 8.1 mm
Wire stripping length: 12 mm
Strain Relief Bracket
You must install the strain relief bracket to ensure that the connector plug does not detach from the instrument during
operation.
Fasten the power cable to the strain relief bracket as shown below, using the supplied tie wraps. This reduces weight
and movement of the AC mains cable at the connector plug.
Make sure the connector
plug is fully inserted.
Install the strain relief
bracket as shown, using
the four hex nuts.
Fasten the AC mains cable
to the strain relief using
the supplied tie wraps.