Specifications

Manual Updates
Manual Updates
Output Terminal Isolation
The Supplemental Characteristics section of this document describes the present output terminal isolation voltage lim-
its for Version 2 and Version 3 instruments.
Version 2 and Version 3 instruments are clearly labeled as such on the rear label located on the AC input cover.
Previous instrument versions DO NOT have version number identification on their rear labels. If you have an instru-
ment without a version number, then the following isolation voltage limits apply:
Instrument without a version number on the rear label
Model voltage rating
Isolation from ground
Positive terminal Negative terminal
80 V +400 V +/-400 V
200 V +600 V +/-400 V
500 V +900 V +/-400 V
750 V +1000 V +/-400 V
1000 V +1000 V +/-400 V
1500 V +1500 V +/-400 V
152 Agilent N8900 Series Operating and Service Guide