Specifications
Calibration Switches
Calibration Switches
SHOCK HAZARD Only qualified, service-trained personnel who are aware of the haz-
ards involved should remove instrument covers. Always disconnect the power cable
and any external circuits before removing the instrument cover.Some circuits are act-
ive and have power for a short time even when the power switch is turned off.
Two switches control the access to calibration commands. The switches are on the interface board located at the back
of the unit and are accessible by removing the top cover. To access the calibration switches:
Accessing the Calibration Switch
1. Remove the instrument cover as described under Disassembly.
2. The calibration switch is on the top of the interface board. To change the calibration switch settings, use a small
screwdriver to move the switches.
3. Replace the top cover when finished.
Do not use a pencil to move the switches. Any graphite dust that gets on the switches will conduct
electricity.
Switches shown
in ON position
Switch Functions
Switches 1 and 2 set the calibration configuration as follows
Switch 1 Switch 2 Description
Normal ON ON
This is the default or as-shipped switch setting. The cal-
ibration functions are accessible after entering a numeric
password. The default password is 0 (zero).
Clear Password OFF ON
The admin/calibration password is reset to 0 when the
instrument is first powered on. Use this setting if you have
forgotten the password.
Inhibit Calibration OFF OFF
All calibration commands are disabled. This is useful where
calibration access is guarded by instrument seals.
Agilent N8900 Series Operating and Service Guide 149