Specifications

Test Record Forms
N8934A/N8954A
N8934A/N8954A Test Record Report Number ______ Date__________
Test Description Model Min. Specs Results Max. Specs
With Remote Sensing
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
24.5 V
Vdmm 500 mV
474.5 V
Vdmm 500 mV
_______
_______
_______
_______
25.5 V
Vdmm + 500 mV
475.5 V
Vdmm + 500 mV
With Local Sensing
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
24.5 V
Vdmm 500 mV
474.5 V
Vdmm 500 mV
_______
_______
_______
_______
25.5 V
Vdmm + 500 mV
475.5 V
Vdmm + 500 mV
CV Load Effect: Both 250 mV _______ + 250 mV
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
+ 350 mV
+ 70 mV
Transient response @ 1.5 ms: Both 5 V _______ + 5 V
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
Both
Both
Both
Both
4.32 A
Idmm 180 mA
85.32 A
Idmm 180 mA
_______
_______
_______
_______
4.68 A
Idmm + 180 mA
85.68 A
Idmm + 180 mA
CC Load Effect: Both 135 mA _______ + 135 mA
Test Description Electronic Load Setting Instrument Setting
Voltage Programming & Readback Min Voltage:
Voltage Programming & Readback High Voltage:
CV Load Effect:
CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min Current:
Current Programming & Readback High Current:
CC Load Effect:
CC, 9 A
CC, 9 A
CC, 0.9 A to 85.5 A
CC, 85.5 A
CC, 9 A to 81 A
CV, 10 V
CV, 10 V
CV, 10 V to 475 V
25 V, 90 A
475 V, 90 A
125 V, 90 A
125 V, 90 A
125 V, 90 A
15 V, 4.5 A
15 V, 85.5 A
500 V, 9 A
Agilent N8900 Series Operating and Service Guide 143