Specifications

Test Record Forms
N8932A/N8952A
N8932A/N8952A Test Record Report Number ______ Date__________
Test Description Model Min. Specs Results Max. Specs
With Remote Sensing
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
9.8 V
Vdmm 200 mV
189.8 V
Vdmm 200 mV
_______
_______
_______
_______
10.2 V
Vdmm + 200 mV
190.2 V
Vdmm + 200 mV
With Local Sensing
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
9.8 V
Vdmm 200 mV
189.8 V
Vdmm 200 mV
_______
_______
_______
_______
10.2 V
Vdmm + 200 mV
190.2 V
Vdmm + 200 mV
CV Load Effect: Both 100 mV _______ + 100 mV
CV Ripple and Noise
peak-to-peak:
rms:
N8932A
N8952A
Both
N/A
N/A
N/A
_______
_______
_______
+ 375 mV
+ 300 mV
+ 40 mV
Transient response @ 1.5 ms: Both 2 V _______ + 2 V
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
Both
Both
Both
Both
10 08 A
Idmm 420 mA
199.08 A
Idmm 420 mA
_______
_______
_______
_______
10 92 A
Idmm + 420 mA
199.92 A
Idmm + 420 mA
CC Load Effect: Both 315 mA _______ + 315 mA
Test Description Electronic Load Setting Instrument Setting
Voltage Programming & Readback Min Voltage:
Voltage Programming & Readback High Voltage:
CV Load Effect:
CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min Current:
Current Programming & Readback High Current:
CC Load Effect:
CC, 21 A
CC, 21 A
CC, 2.1 A to 199.5 A
CC, 199.5 A
CC, 21 A to 189 A
CV, 4 V
CV, 4 V
CV, 4 V to 190 V
10 V, 210 A
190 V, 210 A
50 V, 210 A
50 V, 210 A
50 V, 210 A
9 V, 10.5 A
9 V, 199.5 A
200 V, 21 A
142 Agilent N8900 Series Operating and Service Guide