Specifications
Test Record Forms
N8925A/N8945A
N8925A/N8945A Test Record Report Number ______ Date__________
Test Description Model Min. Specs Results Max. Specs
With Remote Sensing
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
3.92 V
Vdmm – 80 mV
75.92 V
Vdmm – 80 mV
_______
_______
_______
_______
4.08 V
Vdmm + 80 mV
76.08 V
Vdmm + 80 mV
With Local Sensing
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
3.92 V
Vdmm – 80 mV
75.92 V
Vdmm – 80 mV
_______
_______
_______
_______
4.08 V
Vdmm + 80 mV
76.08 V
Vdmm + 80 mV
CV Load Effect: Both – 40 mV _______ + 40 mV
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
+ 320 mV
+ 25 mV
Transient response @ 1.5 ms: Both – 800 mV _______ + 800 mV
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
Both
Both
Both
Both
16.32 A
Idmm – 680 mA
322.32 A
Idmm – 680 mA
_______
_______
_______
_______
17.68 A
Idmm + 680 mA
323.68 A
Idmm + 680 mA
CC Load Effect: Both – 510 mA _______ + 510 mA
Test Description Electronic Load Setting Instrument Setting
Voltage Programming & Readback Min Voltage:
Voltage Programming & Readback High Voltage:
CV Load Effect:
CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min Current:
Current Programming & Readback High Current:
CC Load Effect:
CC, 34 A
CC, 34 A
CC, 3.4 A to 323 A
CC, 323 A
CC, 34 A to 306 A
CV, 3 V
CV, 3 V
CV, 3 V to 76 V
4 V, 340 A
76 V, 340 A
20 V, 340 A
20 V, 340 A
20 V, 340 A
8 V, 17 A
8 V, 323 A
80 V, 34 A
136 Agilent N8900 Series Operating and Service Guide