Specifications
Test Record Forms
N8923A/N8943A
N8923A/N8943A Test Record Report Number ______ Date__________
Test Description Model Min. Specs Results Max. Specs
With Remote Sensing
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
24.5 V
Vdmm – 500 mV
474.5 V
Vdmm – 500 mV
_______
_______
_______
_______
25.5 V
Vdmm + 500 mV
475.5 V
Vdmm + 500 mV
With Local Sensing
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
24.5 V
Vdmm – 500 mV
474.5 V
Vdmm – 500 mV
_______
_______
_______
_______
25.5 V
Vdmm + 500 mV
475.5 V
Vdmm + 500 mV
CV Load Effect: Both – 250 mV _______ + 250 mV
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
+ 350 mV
+ 70 mV
Transient response @ 1.5 ms: Both – 5 V _______ + 5 V
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
Both
Both
Both
Both
1.44 A
Idmm – 60 mA
28.44 A
Idmm – 60 mA
_______
_______
_______
_______
1.56 A
Idmm + 60 mA
28.56 A
Idmm + 60 mA
CC Load Effect: Both – 45 mA _______ + 45 mA
Test Description Electronic Load Setting Instrument Setting
Voltage Programming & Readback Min Voltage:
Voltage Programming & Readback High Voltage:
CV Load Effect:
CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min Current:
Current Programming & Readback High Current:
CC Load Effect:
CC, 3 A
CC, 3 A
CC, 0.3 A to 28.5 A
CC, 28.5 A
CC, 3 A to 27 A
CV, 10 V
CV, 10 V
CV, 10 V to 475 V
25 V, 30 A
475 V, 30 A
125 V, 30 A
125 V, 30 A
125 V, 30 A
15 V, 1.5 A
15 V, 28.5 A
500 V, 3 A
134 Agilent N8900 Series Operating and Service Guide