Specifications

Test Record Forms
N8921A/N8941A
N8921A/N8941A Test Record Report Number ______ Date__________
Test Description Model Min. Specs Results Max. Specs
With Remote Sensing
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
9.8 V
Vdmm 200 mV
189.8 V
Vdmm 200 mV
_______
_______
_______
_______
10.2 V
Vdmm + 200 mV
190.2 V
Vdmm + 200 mV
With Local Sensing
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
9.8 V
Vdmm 200 mV
189.8 V
Vdmm 200 mV
_______
_______
_______
_______
10.2 V
Vdmm + 200 mV
190.2 V
Vdmm + 200 mV
CV Load Effect: Both 100 mV _______ + 100 mV
CV Ripple and Noise
peak-to-peak:
rms:
N8921A
N8941A
Both
N/A
N/A
N/A
_______
_______
_______
+ 375 mV
+ 300 mV
+ 40 mV
Transient response @ 1.5 ms: Both 2 V _______ + 2 V
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
Both
Both
Both
Both
3 36 A
Idmm 140 mA
66.36 A
Idmm 140 mA
_______
_______
_______
_______
3 64 A
Idmm + 140 mA
66.64 A
Idmm + 140 mA
CC Load Effect: Both 105 mA _______ + 105 mA
Test Description Electronic Load Setting Instrument Setting
Voltage Programming & Readback Min Voltage:
Voltage Programming & Readback High Voltage:
CV Load Effect:
CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min Current:
Current Programming & Readback High Current:
CC Load Effect:
CC, 7 A
CC, 7 A
CC, 0.7 A to 66.5 A
CC, 66.5 A
CC, 7 A to 63 A
CV, 4 V
CV, 4 V
CV, 4 V to 190 V
10 V, 70 A
190 V, 70 A
50 V, 70 A
50 V, 70 A
50 V, 70 A
9 V, 3.5 A
9 V, 66.5 A
200 V, 7 A
Agilent N8900 Series Operating and Service Guide 133