Specifications

Calibration Procedure
Step 6. Select the fourth current calibration point. Wait 3 minutes for the internal shunt temperature to stabilize. Cal-
culate the shunt current (I=V/R) and enter the data.
Front Panel SCPI Command
Display shows: Enter P4 measured data”
Enter the data from the external DMM.
This should be the full-scale current rat-
ing. Press Enter when done. Press Back
repeatedly to back out of the calibration
menu.
CAL:LEV P4
*OPC?
CAL:DATA <data>
Step 7. Save the calibration data.
Front Panel Menu SCPI Command
Select System\Admin\CAL\SAVE.
Select Save to save the calibration data.
To save calibration data:
CAL:SAVE
Enter a Calibration Date and Log Out
Front Panel Menu SCPI Command
Select System\Admin\CAL\DATE.
Enter the calibration date in the Date field.
Select System\Admin\Logout
to exit calibration mode.
To enter a calibration date:
CAL:DATE "<date>"
To exit calibration mode:
CAL:STAT OFF
Agilent N8900 Series Operating and Service Guide 131