Specifications
Calibration Procedure
Calibration Procedure
Enter Calibration Mode
Voltage Calibration - With Remote Sensing
Voltage Calibration - With Local Sensing
Current Calibration
Enter a Calibration Date and Log Out
HAZARDOUS VOLTAGES The output may be enabled at turn-on, depending on the
*PON setting. Check this setting before attaching anything to the output terminals.
Introduction
The instrument features closed-case electronic calibration; no internal mechanical adjustments are required. The
instrument calculates correction factors based on input reference values that you set and stores correction factors in
non-volatile memory until the next calibration adjustment is performed. This EEPROM calibration memory is not
changed by cycling power or *RST.
l Refer to Recommended Test Equipment and Setups for the equipment list and test setups required for calibration.
l Refer to the Measurement Techniques section for information about setting up the voltmeter, connecting the cur-
rent shunt, and connecting the electronic load..
l The correct password is required to enter the Admin menu, which contains the calibration function. The password
is factory-set to 0 (zero). You can change the password once calibration mode is entered to prevent unauthorized
access to the calibration mode. Refer to Password Protection for more information.
l When calibrating the unit using SCPI commands, most steps involve sending a *OPC? query to synchronize with
the instrument’s command completion before proceeding. The response from the instrument must be read each
time *OPC? is given. Additionally, the relatively slow settling times and slew rates of the power supply must be
taken into account. Use "Wait" statements to allow for the appropriate settling times for the power supply.
l Any calibration section can be run and saved independently of the other sections. As each calibration section is
completed, the instrument calculates new calibration constants and begins using them. However, these constants
are not saved in nonvolatile memory until a SAVE command is explicitly given.
l Exit the calibration mode either by logging out of the Admin menu or by sending CAL:STAT OFF. Note that any cal-
ibration section that was calibrated but not saved will revert to its previous calibration constants.
HAZARDOUS VOLTAGES All models generate voltages in excess of 60 VDC, with some
models rated at up to 1,500 VDC! Ensure that all instrument connections, load wiring,
and load connections are either insulated or covered so that no accidental contact
with lethal output voltages can occur.
Always turn off the output when connecting or disconnecting any equipment on the
sense or output terminals of the unit.
Agilent N8900 Series Operating and Service Guide 127