Specifications
Recommended Test Equipment and Setups
Recommended Test Equipment and Setups
Test Equipment
Verification and Calibration Setups
Test Equipment
The test equipment recommended for the performance verification and adjustment procedures is listed below. If the
exact instrument is not available, substitute calibration standards of equivalent accuracy.
Instrument Requirements Recommended Model Use
1
Digital Multimeter
Resolution: 10 nV @ 1V;
Readout: 8 1/2 digits
Accuracy: 20 ppm
Agilent 3458A V, C
Current Shunt
2
15 A (0.1Ω)
50 A (0.05Ω)
300 A (0.001Ω)
1000 A (0.0001Ω)
Guildline 9230-15R
Guildline 9230-50, OPT 92310
Guildline 9230-300, OPT 92310
Guildline 9230A/1000, OPT 92310
V, C
Electronic load
2
160 V, 300 A 7.2 kW or equivalent
400 V, 150 A 7.2 kW or equivalent
750 V, 75 A 7.2 kW or equivalent
2 - EA-EL 9160-300 HP
2 - EA EL 9400-150 HP
2 - EA-EL 9750-75 HP
V, C
GPIB controller Full GPIB capabilities Agilent 82350B or equivalent V, C
Oscilloscope
Sensitivity: 1 mV
Bandwidth Limit: 20 MHz
Agilent DSO6054A or equivalent V
RMS Voltmeter
True RMS
Bandwidth: 300 kHz
Agilent 34410A or equivalent V
Voltage divider
5 kV to 5 V (1000 to 1)
Accuracy: 0.025%
Ohms-Lab KVVB-5-5 or equivalent V, C
Disconnect switch 2 pole, 60 A
Square D QO200TR series G03 or
equivalent
V, C
Differential Amplifier Bandwidth: 20 MHz
LeCroy DA1855A, DA1850A, or
equivalent
V
Differential Probe 100:1 / 10:1 selectable LeCroy DXC100A or equivalent V
Terminations 1 – 50Ω BNC termination V
1
V = Verification C = Calibration
2
Dependent on rating of model being tested
Agilent N8900 Series Operating and Service Guide 121