Specifications
Verification and Calibration
Verification and Calibration
Further Verification and Calibration topics:
Recommended Test Equipment and Setups
Performance Verification
Calibration Procedure
Test Record Forms
Verification
The verification procedures verify that the Agilent N8900 series power supply is operating normally and is within pub-
lished specifications. If the instrument fails any of the tests or if abnormal test results are obtained, try calibrating the
unit. If calibration is unsuccessful, return the unit to an Agilent Technologies Service Center.
Agilent Technologies Calibration Services - Your local Agilent Technologies Service Center offers low-cost recal-
ibration. The service center uses automated calibration systems that allow Agilent to provide calibration at competitive
prices.
Agilent Technologies recommends that complete verification should always be performed at the calibration interval. If
the power supply passes the verification tests, the unit is operating within its calibration limits and does not need to be
re-calibrated. This ensures that the instrument will remain within specifications for the next calibration interval and
provides the best long-term stability. Performance data measured using this method can be used to extend future cal-
ibration intervals.
Calibration Interval
The instrument should be calibrated on a regular interval determined by the accuracy requirements of your application.
A 1-year interval is adequate for most applications. Accuracy specifications are warranted only if adjustment is made at
regular calibration intervals. Accuracy specifications are not warranted beyond the 1-year calibration interval.
Test Considerations
For optimum performance, all verification and calibration procedures should comply with the following recom-
mendations:
l Ambient temperature is stable, at 23°C ±5°C.
l Ambient relative humidity is less than 70%.
l 30 minute warm-up period before verification or adjustment.
l Cables as short as possible, twisted or shielded to reduce noise.
Measurement Techniques
Voltmeter
To ensure that the values read by the voltmeter during both the verification procedure and the calibration procedure
are not affected by the instantaneous measurement of the AC peaks of the output current ripple, make several DC
measurements and average them. You can set up the voltmeter to do this automatically by programming ≥10 power
Agilent N8900 Series Operating and Service Guide 119