Specifications
Command Quick Reference
Command Quick Reference
Some [optional] commands have been included for clarity. All settings commands have a corresponding query. See the
Syntax Conventions for SCPI.
ABORT Command
ABORt
:TRANsient Cancels any triggered actions.
CALibrate Commands
CALibrate
:COUNt? Returns the number of times the unit has been calibrated.
:CURRent
[:LEVel] [<value>] Calibrates the current programming and measurement.
:DATA <value> Enters the calibration value read by the external meter.
:DATE <"date"> Enters the calibration date in nonvolatile memory.
:LEVel P1|P2|P3|P4 Advances to the next level in the calibration.
:PASSword <value> Sets a numeric password to prevent unauthorized calibration.
:SAVE Saves the calibration constants in non-volatile memory.
:STATe 0|OFF|1|ON Enables or disables calibration mode.
:VOLTage
[:LOCal]
[:LEVel] [<value>] Calibrates the local voltage programming and measurement.
:REMote
[:LEVel] [<value>] Calibrates the remote voltage programming and measurement.
DISPlay Command
DISPlay
[:WINDow]
[:STATe] 0|OFF|1|ON Turns the front panel display on or off.
IEEE-488 Commands
*CLS Clear status
*ESE<value> Standard event status enable
*ESR? Return event status register
*IDN? Return instrument identification
*LRN? Returns a sequence of SCPI commands
*OPC Enable "operation complete" bit in ESR
*OPT? Return option number
Agilent N8900 Series Operating and Service Guide 105