Product specifications

The Agilent N6700 low-profile
modular power system (MPS) is
a switching regulated, multiple-
output programmable DC power
supply system with the perfor-
mance of a linear power supply.
The N6700 is a flexible modular
platform that allows you to mix
and match more than 30 different
DC power modules to create a
1- to 4-channel DC power system
to optimized performance, power
and price to match test needs.
Test system engineers can invest
in high-performance outputs where
speed and accuracy are needed,
or purchase basic performance
outputs for simple DC power
requirements.
Small size
The Agilent N6700 MPS uses an
advanced switching power supply
design that fits within 1U of rack
space. It has side air vents (no
top or bottom air vents) so other
instruments can be mounted directly
above or below it. (Requires rack
mount kit).
Protection features
Each N6700 module is protected
against over-voltage, over-current,
and over-temperature. A fault con-
dition in one module can be detected
within 10 microseconds by other
modules so that they can be quickly
shut down to avoid hazardous
conditions on your DUT.
High Performance DC Power Supplies speed and accuracy for test optimization
Spe cif icati ons
(at 0° to 55°C, and
derated above 40°C)
N6751A N6752A N6753A N6754A N6755A N6756A
M
ore detailed specifications at
www.agilent.com/find/N6700
75
Small size: up to 4 outputs in 1U of rack space
More than 30 DC power modules: basic, performance and precision models
Fast output programming with active downprogramming
Ultra fast command processing time
Output sequencing and advanced triggering system
Optional LIST mode, built-in digitizer, and disconnect and polarity reversal relays
N6700B, N6701A, N6702A, N6731-36B,
N6741B-46B, N6751A-N6754A,
N6761A-62A, N6773A-N6777A
DC output ratings
Voltage 50 V 50 V 20 V 60 V 20 V 60 V
Current
1
5 A 10 A 50 A 20 A 50 A 17 A
Power 50 W 100 W 300 W 300 W 500 W 500 W
Programming accuracy (at 23°C ±5°C
after 30 minute warm-up. Applies fro
m
minimum to maximum programming range)
Voltage 0.06% + 19 mV 19 mV 10 mV 25 mV 10 mV 25 mV
Current 0.1% + 20 mA 20 mA 30 mA 12 mA 30 mA 12 mA
Voltmeter/ammeter
measurement
accuracy
(at 23°C ±5°C)
Voltage 0.05% + 20 mV 20 mV 10 mV 25 mV 10 mV 25 mV
Current 0.1% + 4 mA 4 mA 30 mA 8 mA 30 mA 8 mA
Output ripple and noise (PARD)
(from 20 Hz to 20 MHz)
CV peak-to-peak 4.5 mV 4.5 mV 5 mV 6 mV 5 mV 6 mV
CV rms 350 µV 350 µV 1 mV 1 mV 1 mV 1 mV
Load effe
c
t (Regulation)
(Applies for any output load change, with a maximum load-lead drop of 1 V/lead.
The load lead drop reduces the maximum available voltage at the load.)
Voltage 2 mV 2 mV 2 mV 2 mV 2 mV 2 mV
Current 2 mA 2 mA 12 mA 5 mA 12 mA 5 mA
Source effect (Regulation)
Voltage 1 mV 1 mV 0.5 mV 1.2 mV 0.5 mV 1.2 mV
Current 1 mA 1 mA 5 mA 2 mA 5 mA 2 mA
Load transient recovery time
(Time to recover to within the settling band following a load change.)
- from 60% to 100% and from 100% to 60% of full load for model N6751A
- from 50% to 100% and from 100% to 50% of full load for models N6752A though N6756A
Voltage settling band
±
75 mV
±
75 mV
2
±
30 mV
3
±
90 mV
4
±
30 mV
3
±
90 mV
4
Time <100 µs <100 µs <100 µs <100 µs <100 µs <100 µs
1
Output current is derated 1% per
°C above 40°C
.
2
When relay option 761 is installed in Model N6752A, the settling band is
±125 mV
.
3
When relay option 760 or 761 is installed on Model N6753A and N6755A, the settling band is
±200 mV
.
4
When relay option 760 or 761 is installed on Model N6754A and N6756A, the settling band is
±350 mV
.
Low-Profile Modular Power System
20-300 W GPIB, LAN, USB, LXI Class C