Product specifications

70
Supplemental characteristics
for all model numbers
DC floating voltage: Output terminal can
be floated up to ±60 VDC from chassis
ground
Remote sensing: Up to half the rated
output voltage can be dropped in each
load lead. The drop in the load leads
subtracts from the voltage available
at the load.
Command processing time: Average
time required for the output voltage
to begin to change following receipt of
digital data is 20 ms for power supplies
connected directly to the GPIB.
Modulation: (analog programming of
output voltage and current):
Input signal: 0 to -5 V for voltage,
and 0 to +5 V for current.
Input impedance: 30 kΩ or greater.
AC input (47 to 63 Hz): 180 to 235 VAC
(line-to-line 3 phase) 36 Arms maximum
worst case, 28 Arms nominal;
360 to 440 VACX, 18 Arms maximum
worst case, 14 Arms nominal. (Maximum
line current includes 5% unbalanced
phase voltage condition).
Software driver:
•IVI-COM
•VXIPlug&Play
Input power: 9000 VA and 7950 W
maximum; 175 W at no load.
Size: 425.5 mm W x 221.5 mm H x
674.7 mm D (16.75 in x
8.75 in x 25.56 in).
Warranty: One year
Ordering information
Opt 208 180 to 235 VAC, 3 phase,
47 to 63 Hz
Opt 400 360 to 440 VAC, 3 phase,
47 to 63 Hz
Opt 602 Two bus bar spacers for
paralleling power supplies
(p/n 5060-3514)
* Opt 908 Rack-mount kit (p/n 5062-3977
and p/n 5063-9212)
* Opt 909 Rack-mount kit with handles
(p/n 5063-9221 and p/n 5063-9219).
Single-Output: 6600 W GPIB (Continued)
More detailed specifications at
www.agilent.com/find/6690
High Performance DC Power Supplies speed and accuracy for test optimization
Spe cif icati ons
(at 0° to 55°C unless
otherwise specified)
6690A 6691A 6692A
Supplemental Characteristics (Non-warranted characteristics determined by design that are
useful in applying this product)
Ripple and noise constant current mode
from 20 Hz to 20 MHz
rms 200 mA 50 mA 30 mA
Average programming resolution
Voltage 4.1 mV 8.1 mV 16 mV
Current 118.5 mA 59 mA 30 mA
OVP 90 mV 170 mV 330 mV
Output voltage programming
response time (excludes 45 ms 60 ms 100 ms
command-processing time)
Full-load progrmming rise or fall time
(10 to 90% or 90 to 10%, resistive load)
Output common-mode
noise current rms 3 mA 3.5 mA 4 mA
(to signal-ground peak-to-peak 20 mA 20 mA 25 mA
binding post)
Opt 0L1 Full documentation on
CD-ROM, and printed standard
documentation package
Opt 0L2 Extra copy of standard
printed documentation package
Opt 0B0 Full documentation on
CD-ROM only
Opt 0B3 Service manual
* Support rails required
Accessories
p/n 5065-6935 Replacement fuse kit for
360-440 VAC line.
p/n 5065-6934 Replacement fuse kit for
180-235 VAC line.
E3663AC Support rails for Agilent
rack cabinets.
p/n 5080-2148 Serial link cable 2 m
(6.6 ft.)
+
–
Side
Output
Buss Bars
Top
Buss Bar Detail
Rear
12.7 mm
0.5"
28.6 mm
1.13"
30.2 mm
1.19"
65.1 mm
2.56"
27.8 mm
1.09"
9.5 mm
0.38"
12.7 mm
0.5"
674.7 mm
26.56"
667.3 mm
26.27"
574 mm
22.6"
221.5 mm
8.75"
425.5 mm
16.75"
3/8" Diameter (6)
3 3/4"
+
–
Side
Output
Buss Bars
Top
Buss Bar Detail
Rear
12.7 mm
0.5"
28.6 mm
1.13"
30.2 mm
1.19"
65.1 mm
2.56"
27.8 mm
1.09"
9.5 mm
0.38"
12.7 mm
0.5"
674.7 mm
26.56"
667.3 mm
26.27"
574 mm
22.6"
221.5 mm
8.75"
425.5 mm
16.75"
3/8" Diameter (6)
3 3/4"
Agilent Models: 6690A, 6691A, 6692A