Product specifications

35
Single-Output 3.3-5.2 KW GPIB, LAN, USB
(Continued)
Spe cif icati ons
(at 0° to 55°C unless
otherwise specified)
N8754A N8755A N8756A N8757A N8758A
Number of outputs
1 1 1 1 1
GPIB
Yes Yes Yes Yes Yes
Output ratings
Voltage
1
20 V 30 V 40 V 60 V 80 V
Current
2
250 A 170 A 125 A 85 A 65 A
Power
5000 W 5100 W 5000 W 5100 W 5200 W
Programming accuracy
Voltage
1
0.025% + 15 mV 22.5 mV 30 mV 45 mV 60 mV
Current
2, 6
0.1% + 750 mA 510 mA 375 mA 255 mA 195 mA
Output ripple and noise
CV p-p (Up to 20 MHz) 75 mV 75 mV 75 mV 75 mV 100 mV
CV rms (From 5 Hz – 1 MHz) 10 mV 10 mV 10 mV 10 mV 15 mV
Measurement accuracy
Voltage 0.025% + 25 mV 37.5 mV 50 mV 75 mV 100 mV
Current
6
0.15% + 750 mA 510 mA 375 mA 255 mA 195 mA
Load effect
CV load regulation
3
8 mV 9.5 mV 11 mV 14 mV 17 mV
CC load regulation
4
250 mA 170 mA 125 mA 85 mA 65 mA
Source effect
CV line regulation
5
2 mV 3 mV 4 mV 6 mV 8 mV
CC line regulation
5
125 mA 85 mA 62.5 mA 42.5 mA 32.5 mA
Load transient recovery time
Time
7
<1 ms <1 ms <1 ms <1 ms <1 ms
Supplemental Characteristics
(Non-warranted characteristics determined by design and
useful in applying the product)
Output response time
Up-prog response time
8
30 ms 30 ms 30 ms 50 ms 50 ms
Down-prog response time Full-load
8
50 ms 80 ms 80 ms 80 ms 100 ms
Down-prog response time No-load
9
700 ms 800 ms 900 ms 1000 ms 1200 ms
Remote sense compensation 2 V 5 V 5 V 5 V 5 V
Output ripple and noise
CC rms
10
1000 mA 460 mA 300 mA 150 mA 120 mA
Programming resolution
Measurement resolution
Voltage 2.4 mV 3.6 mV 4.8 mV 7.2 mV 9.6 mV
Current 30 mA 20.4 mA 15 mA 10.2 mA 7.8 mA
More detailed specifications at
www.agilent.com/find/N8700
Basic DC Power Supplies essential features for a tight budget
Supplemental characteristics
for all model numbers
Command response time: Average time
required for the output voltage to begin
to change following receipt of digital
data is 100 ms.
Output terminal isolation:
8 V to 60 V units
No output terminal may be more than
±60 VDC from any other terminal or
chassis ground
80 V to 600 V units
No positive output terminal may be more
than ± 600 VDC from any other terminal
or chassis ground.
No negative output terminal may be
more than ± 400 VDC from any other
terminal or chassis ground.
Modulation: (Analog programming
of output voltage and current)
Input signal: selectable,
0 to 5 V/0 to 10 V full scale
Input impedance: selectable,
0 to 5 kΩ/0 to 10 kΩ full scale
I/O interface: GPIB, LAN, USB standard
Software driver:
• IVI-COM
• LabVIEW
AC input:
Input range:
230 VAC single-phase option
170 - 265 VAC; 47 - 63 Hz;
3.3 kW models only
208 VAC 3-phase option
170 - 265 VAC; 47 - 63 Hz
400 VAC 3-phase option
342 - 460 VAC; 47 - 63 Hz
Input current:
230 VAC single-phase
3.3 kW models only: 23 - 24 A Max @ 100% load
208 VAC 3-phase option
3.3 kW models: 13.6 - 14.5 A Max @ 100% load
5 kW models: 21-22 A max @ 100% load
400 VAC 3-phase option
3.3 kW models: 6.8 - 7.2 A Max @ 100% load
5 kW models: 10.5 - 12 A Max @ 100% load
Notes:
1
Minimum voltage is guaranteed to maximum
0.2% of rated output voltage.
2
Minimum current is guaranteed to maximum
0.4% of rated output current.
3
From no-load to full-load, constant input voltage.
Maximum drop in remote sense.
4
For load voltage change equal to the unit voltage
rating, constant input voltage
5
Single-phase and 3-Phase 208 V models:
170~265 VAC, constant load.
3-Phase 400 V models: 342~460 VAC, constant load.
6
The constant current programming readback
and monitoring accuracy does not include the
warm-up and load regulation thermal drift.
7
Time for output voltage to recover within 0.5%
of its rated output for a load change 10 - 90%
of rated output current, local sense.
8
From 10% to 90% or 90% to 10% of rated output
voltage, with rated, resistive load.
9
From 90% to 10% of rated output voltage.
10
For 8 V - 15 V models the ripple is measured from
2 V to rated output voltage and rated output current.
For other models, the ripple is measured at 10 - 100%
of rated output voltage and rated output current.