Product specifications

This is the DC power supply for
every engineer’s or electronic
technician’s lab bench. It has two
tracking 25 V outputs, which are
together referenced to a floating
common, and an isolated 6 volt
output. It is easy to control from
the front panel, or with industry
standard SCPI commands via the
GPIB or RS232. VXIPlug&Play
drivers are available to further
simplify computer control. Up to
3 complete states can be stored for
later recall. The low noise, excellent
regulation, and built-in voltmeter/
ammeter make this reliable power
supply well suited for the needs
of the R&D lab.
Basic DC Power Supplies essential features for a tight budget
Spe cif icati ons
(at 0° to 55°C unless
otherwise specified)
E3631A
More detailed specifications at
www.agilent.com/find/E3600
19
Small, compact size for bench use
Low output ripple and noise
Built-in measurements and basic programmable features
Over-voltage protection to ensure DUT safety
Three memory states to ease test setup
E3631A
Triple-Output
80 W GPIB
DC outputs
Voltage 0 to +25 V 0 to -25 V 0 to 6 V
Current 0 to 1 A 0 to 1 A 0 to 5 A
Load regulation
Voltage <0.01% + 2 mV <0.01% + 2 mV <0.01% + 2 mV
Current <0.01% + 250 µA <0.01% + 250 µA <0.01% + 250 µA
Line regulation
Voltage <0.01% + 2 mV <0.01% + 2 mV <0.01% + 2 mV
Current <0.01% + 250 µA <0.01% + 250 µA <0.01% + 250 µA
Ripple and noise
from 20 Hz to 20 MHz
Normal-mode voltage <350 µVrms/2 mVp-p <350 µVrms/2 mVp-p <350 µVrms/2 mVp-p
Normal-mode current <500 µArms <500 µArms <2 mArms
Common-mode current <1.5 µArms <1.5 µArms <1.5 µArms
Programming accuracy
at 25°C ±5°C
Voltage 0.05% + 20 mV 0.05% + 20 mV 0.1% + 5 mV
Current 0.15% + 4 mA 0.15% + 4 mA 0.2% + 10 mA
Readback accuracy at 25°C ±5°C
Voltage 0.05% + 10 mV 0.05% + 10 mV 0.1% + 5 mV
Current 0.15% + 4 mA 0.15% + 4 mA 0.2% + 10 mA
Resolution
Program/readback 1.5 mV, 0.1 mA 1.5 mV, 0.1 mA 0.5 mV, 0.5 mA
Meter 10 mV, 1 mA 10 mV, 1 mA 1 mV, 1 mA
Transient response 50 µs for output to recover to within 15 mV following a change in output
current from full load to half load or vice versa
For off-the-shelf shipment
Application notes
Understanding Linear Power
Supply Operation (AN1554)
5989-2291EN
10 Practical Tips You Need to
Know About Your Power Products
5965-8239E
Supplemental characteristics
for all model numbers
Product regulation: Designed to
comply with UL1244, IEC 1010-1;
certified with CSA 22.2
Meets requirements for CE regulation
Software driver:
IVI-COM
VXIPlug&Play
IntuiLink Connectivity Software
Size: E3631A:
213 mm W x 133 mm H x 348 mm D
(8.4 in. x 5.2 in. x 14.2 in.)
Weight: E3631A: 8.2 kg (18 lbs)
Warranty: One year
Ordering information
Opt 0E9 90 to 110 VAC, 47 to 63 Hz
(Japan only)
Opt 0EM 104 to 126 VAC, 47 to 63 Hz
Opt 0E3 207 to 253 VAC, 47 to 63 Hz
Opt 1CM Rack mount kit
Opt 0L2 Extra documentation package