Product specifications

105
Notes:
1. Operating temperature range is
0° to 55°C. All specifications apply
for 25°C ±5°C, except as noted.
2. Maximum continuous power
available is derated linearly from
40°C to 75% of maximum at 55°C.
3. DC current accuracy specifications
apply 30 seconds after input
is applied.
Single-Input: 250 W to 300 W (Continued)
Spe cif icati ons
More detailed specifications at
www.agilent.com/find/6060
DC Electronic Loads maximize throughput with real-life loading conditions
6060B 6063B
Supplemental Characteristics (Non-warranted characteristics determined by design that are
(Continued)
useful in applying the product)
Analog monitor accuracy
Current monitor (0 to 10 V
out
) 4% ±85 mA 3% ±10 mA
Temperature coefficient 50 ppm/°C ±6 mA/°C 100 ppm/°C ±1 mA/°C
Voltage monitor (0 to 10 V
out
) 0.25% ±40 mV 0.4% ±240 mV
Temperature coefficient 50 ppm/°C ±0.2 mV/°C 70 ppm/°C ±1.2 mV/°C
Remote sensing 5-VDC maximum between 5-VDC maximum between
sense and load input sense and load input
Minimum operating voltage 2 volts (1.2 V typical) 2 volts (1.2 V typical)
(at full rated current)
Programmable short 0.033 Ω (0.020 Ω typical) 0.20 Ω (0.10 Ω typical)
Programmable open (typical) 20 kΩ 80 kΩ
Drift (over 8-hour interval)
Current 0.03% ±10 mA 0.03% ±15 mA
Voltage 0.01% ±10 mV 0.01% ±20 mV
DC isolation voltage ±240 VDC, between any ±240 VDC, between any
input and chassis ground input and chassis ground
Digital inputs V
IL
= 0.9 V
max
at I
IL
= V
IL
= 0.9 V
max
at I
IL
=
–1 mA / V
IH
= 3.15 V
min
–1 mA / V
IH
= 3.15 V
min
(pull-up resistor on input) (pull-up resistor on input)
Digital outputs V
OL
= 0.72 V
max
at I
OL
= V
OL
= 0.72 V
max
at I
OL
=
1 mA / V
OH
= 4.4 V
min
1 mA / V
OH
= 4.4 V
min
at I
OH
= –20 µA at I
OH
= –20 µA
Net weight (approx.) 6.12 kg (13.5 lb) 6.12 kg (13.5 lb)