Specifications

Output ratings
Voltage 20 V 35 V 60 V 100 V
Current
1
15 A
2
8.5 A 5 A 3 A
Power 300 W 300 W 300 W 300 W
Programming accuracy (at 23°C ±5°C)
Voltage 0.1% + 0.1% + 0.1% + 0.1% +
20 mV 35 mV 60 mV 100 mV
Current 0.15% + 0.15% + 0.15% + 0.15% +
60 mA 60 mA 60 mA 30 mA
Readback accuracy (at 23°C ±5°C)
Voltage 0.1% + 0.1% + 0.1% + 0.1% +
20 mV 35 mV 60 mV 100 mV
Current 0.15% + 0.15% + 0.15% + 0.15% +
15 mA 12 mA 12 mA 6 mA
Output ripple and noise (PARD) (from 20 Hz to 20 MHz)
CV peak-to-peak 20 mV 22 mV 35 mV 45 mV
CV rms 3 mV 5 mV 9 mV 18 mV
Load regulation
4
Voltage 13 mV 16 mV 24 mV 45 mV
Current 6 mA 6 mA 6 mA 6 mA
Line regulation
Voltage 2 mV 4 mV 6 mV 10 mV
Current 1 mA 1 mA 1 mA 1 mA
Transient response time (time to recover to within the settling band following a load change from 50% to 100%
and from 100% to 50% of full load.)
Voltage settling band ± 0.3 V
3
± 0.3 V
3
± 0.5 V ± 1.0 V
Time <250 µs <250 µs <250 µs <250 µs
Programming resolution
Voltage 7 mV 10 mV 18 mV 28 mV
Current 9 mA 6 mA 3 mA 1.5 mA
Output ripple and noise (PARD)
CC rms 6 mA 6 mA 6 mA 6 mA
Over-voltage protection
Accuracy 0.25% + 0.25% + 0.25% + 0.25% +
100 mV 130 mV 260 mV 650 mV
Accuracy w/opt 760 0.25% + 0.25% + 0.25% + 0.25% +
700 mV 700 mV 400 mV 650 mV
Accuracy w/opt 761 0.25% + 0.25% + 0.25% + 0.25% +
500 mV 350 mV 350 mV 650 mV
Maximum setting 22 V 38.5 V 66 V 110 V
Response Time
50 µs from ocurence of O V condition to start of output shutdown
Maximum up-programming time with full resistive load: (time from 10% to 90% of total voltage excursion)
Voltage setting from 0 V to 10 V 20 ms 20 ms 20 ms 20 ms
Maximum up-programming settling time with full resistive load: (time from start of voltage change to within 50 mV of final value)
Voltage setting from 0 V to 10 V 100 ms 100 ms 100 ms 100 ms
1
Output current is derated 1% per ºC above 40ºC.
2
When relay Option 760 is installed, the maximum output current will be limited to 10 A.
3
When relay Option 760 or 761 is installed, the settling band is ±0.35 V.
4
With output change from no load to full load, up to a maximum load-lead drop of 1 V/lead.
78
Power management feature allows
you to allocate mainframe power
To further optimize your investment
you may choose to save money
configuring a system where the
sum of the power modules installed
in a mainframe exceeds the total
power available from the main-
frame. In this case, the new power
management features of the N6700
allows you to allocate mainframe
power to the outputs where it’s
needed and reduce power to the
outputs where it is not needed,
achieving maximum asset utilization
and flexibility. This feature provides
the safety from unexpected and
dangerous shutdowns that can
occur with power systems without
power management when operated
in a similar way.
Series and parallel operation
To increase the available power,
similarly rated outputs can be
operated in series for greater
output voltage or in parallel for
greater output current.
To simplify parallel operation, the
N6700 offers virtual channels, a
firmware based feature that allows
the N6700 system to treat up to
4 channels as a single, synchronized
channel. Once configured, all func-
tions (sourcing, measurements,
triggering, protection, and status
monitoring) behave as if there is
1 channel of up to 4 times the
capacity of a single channel, with-
out writing a single line of code
to manage the interaction and
synchronization of the paralleled
power supplies.
Low-Profile Modular Power System
20-300 W GPIB (Continued)
M
ore detailed specifications at
www.agilent.com/find/N6700
High Performance DC Power Supplies speed and accuracy for test optimization
Specific ation s
(at 0° to 55°C, and
derated above 40°C)
N6773A N6774A N6775A N6776A