Specifications

The Agilent N6700 low-profile
modular power system (MPS) is
a switching regulated, multiple-
output programmable DC power
supply system with the perfor-
mance of a linear power supply.
The N6700 is a flexible modular
platform that allows you to mix
and match more than 20 different
DC power modules to create a
1- to 4-channel DC power system
to optimized performance, power
and price to match test needs.
Test system engineers can invest
in high-performance outputs where
speed and accuracy are needed,
or purchase basic performance
outputs for simple DC power
requirements.
Small size
The Agilent N6700 MPS uses an
advanced switching power supply
design that fits within 1U of rack
space. It has side air vents (no
top or bottom air vents) so other
instruments can be mounted directly
above or below it. (Requires rack
mount kit).
Protection features
Each N6700 module is protected
against over-voltage, over-current,
and over-temperature. A fault con-
dition in one module can be detected
within 10 microseconds by other
modules so that they can be quickly
shut down to avoid hazardous
conditions on your DUT.
High Performance DC Power Supplies speed and accuracy for test optimization
Specific ation s
(at 0° to 55°C, and
derated above 40°C)
N6751A N6752A N6753A N6754A N6761A N6762A
M
ore detailed specifications at
www.agilent.com/find/N6700
76
Small size: up to 4 outputs in 1U of rack space
More than 20 DC power modules: basic, performance and precision models
Fast output programming with active downprogramming
Ultra fast command processing time
Output sequencing and advanced triggering system
Optional LIST mode, built-in digitizer and disconnect relays
N6780 SMU and application specific modules available (pages 91, 106)
N6700B, N6701A, N6702A, N6731-36B,
N6741B-46B, N6751A-N6754A, N6761A-62A,
N6773A-N6776A, N6781A-84A (page 106)
Output ratings
Voltage 50 V 50 V 20 V 60 V 50 V 50 V
Current 5 A 10 A 50 A 20 A 1.5 A 3 A
Power 50 W 100 W 300 W 300 W 50 W 100 W
Programming accuracy (at 23°C ±5°C)
Voltage high range 0.06% + 0.006% + 0.06% + 0.06% + 0.016% + 0.016% +
19 mV 19 mV 10 mV 25 mV 6 mV 6 mV
Voltage low range (≤ 5.5 V) N/A N/A N/A N/A 0.016% + 0.016% +
1.5 mV 1.5 mV
Current high range 0.1% + 0.1% + 0.1% + 0.1% + 0.04% + 0.04% +
20 mA 20 mA 30 mA 8 mA 200 µA 200 µA
Current low range
≤ 100 mA, @ 0 - 7 V
N/A N/A N/A N/A 0.04% + 0.04% +
15 µA 15 µA
100 mA, @ 0 - 50 V N/A N/A N/A N/A 0.04% + 0.04% +
55 µA 55 µA
≤ 200 µA, @ 0 - 50 V N/A N/A N/A N/A 0.5% + 0.5% +
100 nA 100 nA
Readback accuracy (at 23°C ±5°C)
Voltage high range
0.05% + 0.05% + 0.05% + 0.05% + 0.016% + 0.016% +
20 mV 20 mV 10 mV 25 mV 6 mV 6 mV
Voltage low range
≤ 5.5 V N/A N/A N/A N/A 0.016% + 0.016% +
1.5 mV 1.5 mV
Current high range
0.1% + 0.1% + 0.1% + 0.1% + 0.04% + 0.04% +
4 mA 4 mA 30 mA 8 mA 160 µA 160 µA
Current low range
100 mA, @ 0 - 7 V
1
N/A N/A N/A N/A 0.03% + 0.03% +
15 µA 15 µA
100 mA, @ 0 - 50 V N/A N/A N/A N/A 0.03% + 0.03% +
55 µA 55 µA
Output ripple and noise (PARD)
(from 20 Hz to 20 MHz)
CV peak-to-peak 4.5 mV 4.5 mV 5 mV 6 mV 4.5 mV 4.5 mV
CV rms 350 µV 350 µV 1 mV 1 mV 350 µV 350 µV
Load regulation
Voltage 2 mV 2 mV 2 mV 2 mV 0.5 mV 0.5 mV
Current 2 mA 2 mA 12 mA 5 mA 65 µA 65 µA
Line regulation
Voltage 1 mV 1 mV 0.5 mV 1.2 mV 0.5 mV 0.5 mV
Current 1 mA 1 mA 5 mA 2 mA 30 µA 30 µA
1
Applies when measuring 4006 data points (SENSe:SWEep:POINts = 4096).
Low-Profile Modular Power System
20-300 W GPIB, LAN, USB, LXI Class C