Technical data
30 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20/40 GHz
3 Battery Removal/Replacement
Replacement Procedure
1 To install the A18BT1 3 V, 0.16 A Lithium battery with p/n 1420-0314 (Panasonic
BR 2325), the positive side is aligned with the positive sign on the A18’s battery clip.
2 Apply the tape that was removed or a non-conductive tape such as
Permacel Capcom tape.
3 Reverse the remaining steps of the removal procedure.
4 Perform the post-repair adjustments and performance tests that pertain to this
removal procedure.
CAUTION
Apply tape over the battery to keep it from shaking free during instrument
movement. If the tape is no longer available or does not stick, use non-conductive
tape, but do not use Scotch tape that could cause static build up.