Technical data
Battery Removal/Replacement 3
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20/40 GHz 29
Step 4. A18BT1 Battery Removal/Replacement
Tools Required
• T-10 driver (N8212A only)
• flat-head screw driver
Removal Procedure
1 Lay the A18 CPU board on a static safe work surface.
2 Remove the tape holding down the battery. Save this tape to reapply over the new
battery.
3 Using the flat-head screw driver, remove the A18BT1 by leveraging the battery out of
the battery’s socket.
Figure 5 A18BT1 Battery
WARNING
This battery contains lithium. Do not incinerate or puncture this battery. Do not
install this battery backwards. Dispose of the battery in a safe manner and in
accordance with your country's requirements. You can return batteries to your
nearest Agilent Technologies Sales and Service office for disposal, if required.
Refer to “Lithium Battery Disposal" on page 21.
A18BT1