Technical data

16 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20/40 GHz
1 Security Features
Remarks
You can exit the secure mode by entering :SYST:SECUrity:LEVel NONE, or by
cycling the power.
:SECurity:OVERwrite
:SYSTem:SECurity:OVERwrite
This command removes all user files, table editor files, flatness correction files, and
baseband generator files. The memory is then overwritten with random data as follows:
SRAM All addressable locations will be overwritten with random characters.
HARD DISK All addressable locations will be overwritten with random characters.
FLASH MEMORY The flash blocks will be erased.
:SECurity:SANitize
:SYSTem:SECurity:SANitize
This command removes all user files, table editor files, flatness correction files, and
baseband generator files. The memory is then overwritten with a sequence of data as
follows:
SRAM All addressable locations will be overwritten with random characters.
HARD DISK All addressable locations will be overwritten with a single character
and then a random character.
FLASH MEMORY The flash blocks will be erased.