Technical data

74 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
3 Auto Couple
Auto All
When Auto All is pressed, the N8201A automatically couples all instrument settings that
have auto and manual choices. In most cases, these auto settings give the most accurate
measurements and optimum dynamic range. When a function is coupled, it is in the Auto
state. When it is uncoupled it is in the Man state.
Key Path: Auto Couple
Dependencies/Couplings: When Auto All is selected:
Resolution BW couples to: Span and Span/RBW
Video BW couples to: Res BW and VBW/RBW
Sweep Time couples to: Res BW; Video BW; Detector; Span and Center Frequency
CF Step couples to: Span in swept spans, to Res BW in zero span
Attenuation couples to: Ref Level; Ext Amp Gain; Atten Step; Max Mixer Lvl; and Int
Preamp
FFT & Sweep couples to: Res BW and Span
PhNoise Opt (phase noise optimization) couples to: Res BW; Span and FFT & Sweep
sweep type
Detector couples to: marker functions; Avg/VBW Type; Average On Off; Max Hold and
Min Hold
Average Type couples to: the marker functions; Detector and Scale Type
ADC Dither couples to: Sweep Type; Span; Res BW; ADC Ranging and FFTs/Span
VBW/RBW ratio is set to 1.0
Span/RBW ratio is set to Auto
Auto Sweep Time is set to Normal
FFT & Sweep is set to Auto:Best Dynamic Range
ADC Ranging is set to Autorange
Marker Count, Gate Time is set to Auto
Marker Trace and Printer have an Auto setting, but are not affected by Auto All.
Remote Command:
:COUPle ALL|NONE
:COUPle?
Remote Command Notes:
COUPle NONE sets all the functions to the manual (not coupled) mode. There is no
front-panel key equivalent to the COUPle NONE command.