Technical data
Amplitude / Y Scale 2
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 71
Atten Step
Permits the selection of 2 dB or 10 dB step resolution for input attenuation.
Key Path: AMPLITUDE / Y Scale
Saved State: Saved in Instrument State
Factory Preset: 2 dB
Remote Command:
[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement] <integer>
(in dB)
[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement]?
Example:
POW:ATT:STEP 10
Sets the Attenuation to 10 dB. The only valid suffix is dB. If no suffix is sent, dB is assumed.
If a value >5 is entered, 10 is used.
If a value ≤5 is entered, 2 is used