Technical data
68 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
2 Amplitude / Y Scale
Delete Point
Allows you to delete the amplitude-correction data for the currently selected point. The
prompt, “If you are sure, press key again to delete” will appear on the display. Pressing
Delete Point again will delete the point and adjust all of the point numbers as appropriate.
Key Path: AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User), Edit
Remote Command:
See “Edit" on page 66.
Delete Corrections
Allows you to clear all data from the selected amplitude-correction set. The prompt If
you are sure, press key again to delete will appear on the display. Pressing
Delete again will delete the correction set.
Key Path: AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User)
Remote Command:
[:SENSe]:CORRection:CSET[1]|2|3|4:DELete
Remote Command Notes: CSET number equivalents to front-panel access definitions
are as follows:
• CSET or CSET1 is Antenna
• CSET2 is Cable
• CSET3 is Other
• CSET4 is User
Example:
CORR:CSET4:DEL
Freq Interp
Allows you to determine how trace values are computed between points in a correction
table. If the linear mode is selected, a straight line is used between points in the correction
table (for a linear frequency scale and for a decibel amplitude scale). If the logarithmic mode
is selected, frequency values between points are computed by first taking the logarithm of
both table values and the intermediate value, while using decibel amplitude values.
Key Path: AMPLITUDE / Y Scale, More, Corrections
State Saved: Not saved in instrument state