Technical data
Amplitude / Y Scale 2
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 67
Point
Allows you to create or edit an amplitude-correction factor data point. Up to 200 points may
be defined for each set. Enter the point number to be created or edited by using the
keyboard, then press Enter. After selecting a point, Frequency becomes active.
Key Path: AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User), Edit
State Saved: Not part of instrument state, saved in a corrections file.
Remote Command:
See “Edit" on page 66.
Frequency
Allows you to enter the frequency value for an amplitude-correction point. Enter the
frequency value by using the keyboard.
A frequency coordinate must always be specified for amplitude-correction factors.
Amplitude-correction data is sorted in the table by frequency. The sorting occurs
immediately after you have entered the frequency value via the front-panel.
The amplitude correction entered for the lowest frequency will be applied to all frequencies
less than the lowest frequency entered. Similarly, the amplitude correction for the highest
frequency entered will be applied to all frequencies greater than the highest frequency
entered.
Key Path: AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User), Edit
State Saved: Not part of instrument state, saved in a corrections file.
Remote Command:
See “Edit" on page 66.
Amplitude
Allows you to enter the amplitude value for the current amplitude-correction point. After
entering an amplitude, the point number automatically increments and Frequency becomes
active to allow entry of the frequency of the next point.
Key Path: AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User), Edit
State Saved: Not part of instrument state, saved in a corrections file.
Remote Command:
See “Edit" on page 66.