Technical data
66 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
2 Amplitude / Y Scale
CORR:CSET2?
Edit
Accesses menu keys that allow you to create and edit an amplitude-correction factor set. It
puts the N8201A into a split-screen mode where the correction data is displayed in a table
under the trace data. Pressing ESC while in this menu will exit the menu and remove the
table from the screen. New points will be applied only after the editor is closed.
Key Path: AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User)
Remote Command:
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA
<freq>,<rel_ampl>{,<freq>,<rel_ampl>} Creates an amplitude-correction
factor set
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA:MERGe
<freq>,<rel_ampl>{,<freq>,<rel_ampl>} Adds the points with the
specified values to the current amplitude correction data, allowing you to merge correction
data. If too much data is merged, as many points as possible are merged into the existing
data and then an error is reported.
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA?
Remote Command Notes:
[:SENSe]:CORRection:CSET:ALL[:STATe] must be set to on for this
command to function.
CSET number equivalents to front-panel access definitions are as follows:
• CSET or CSET1 is Antenna
• CSET2 is Cable
• CSET3 is Other
• CSET4 is User
<freq> is the frequency (in Hz) where the correction should be applied; no unit is allowed in
this parameter
<rel_ampl> is the amount of relative amplitude correction (in dB) needed; no unit is allowed
in this parameter
Example:
:CORR:CSET2:DATA 900E6,0.3,1.0E9,0.35,1.3E9,0.2