Technical data

Amplitude / Y Scale 2
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 65
Antenna, Cable, Other, and User Keys
Pressing Antenna, Cable, Other, or User accesses the Correction menu for that type of
correction data. These four keys display the status of correction sets. If the key indicates
On, then amplitude corrections for this type have been enabled. To perform the corrections,
both this key and Apply Corrections must indicate Yes. The status is toggled in the
correction set menu located under Correction.
Key Path: AMPLITUDE / Y Scale, More, Corrections
Remote Command:
There is no remote command for this key.
Correction
Turns the amplitude correction function on or off for the specific set of correction data. The
corrections state must be set to
On for the correction to be applied.
Antenna, Cable, and Other correction factors are generally entered as positive values. This
indicates a loss in the external device. User correction factors are typically entered as
negative values which indicate a gain in the external device.
Key Path: AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User)
State Saved: Saved in Instrument State
Factory Preset: No
Remote Command:
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]?
Remote Command Notes:
[:SENSe]:CORRection:CSET:ALL[:STATe] must be set to on for this
command to function.
CSET number equivalents to front-panel access definitions are as follows:
CSET or CSET1 is Antenna
CSET2 is Cable
CSET3 is Other
CSET4 is User
Example:
CORR:CSET2 ON