Technical data
64 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
2 Amplitude / Y Scale
Corrections
Accesses the Corrections menu keys that allow you to enable the corrections function and
to select which set of correction factors you wish to modify. These frequency and amplitude
corrections will be applied to the displayed data to correct for system losses or gains
outside the N8201A. Four different sets of correction data can be stored.
Key Path: AMPLITUDE / Y Scale, More
Remote Command:
There is no equivalent remote command for this key.
Apply Corrections
Pressing Apply Corrections (Yes) turns on the amplitude-correction factors. Corrections will
only be applied for the sets of correction factors whose correction state is set to On. To turn
a set of correction factors on, use the Correction On Off key in the Antenna, Cable, Other, or
User menus.
Key Path: AMPLITUDE / Y Scale, More, Corrections
Annunciation/Annotation: When Apply Corrections (Yes) is selected, an A will appear
in the screen annotation on the left edge of the display, whether or not a correction set has
been turned on using the Correction (On) key in the Antenna, Cable, Other, or User menus.
State Saved: Saved in Instrument State
Factory Preset: No
Remote Command:
[:SENSe]:CORRection:CSET:ALL[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET:ALL[:STATe]?
Remote Command Notes:
To turn On or Off an individual correction set, use:
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]
Example:
CORR:CSET:ALL ON
CORR:CSET:ALL?