Technical data
60 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
2 Amplitude / Y Scale
Example:
UNIT:POW DBUVM
dBμA/m
Sets the amplitude units to dBμA/m. This is a unit specifically applicable to EMI field
strength measurements. In the absence of a correction factor this unit is treated by the
instrument exactly as though it were dBmV. You must load an appropriate correction factor
using amplitude corrections for this unit to generate meaningful results. Therefore, this key
is unavailable unless one of the corrections is turned on (in Amplitude, Corrections menu)
and Apply Corrections is set to Yes.
Key Path: AMPLITUDE / Y Scale, Y Axis Units
Remote Command:
See“Y Axis Units" on page 56.
Example:
UNIT:POW DBUAM
dBpT
Sets the amplitude units to dBpT. This is a unit specifically applicable to EMI field strength
measurements. In the absence of a correction factor this unit is treated by the instrument
exactly as though it were dBmV. You must load an appropriate correction factor using
amplitude corrections for this unit to generate meaningful results. Therefore, this key is
unavailable unless one of the corrections is turned on (in Amplitude, Corrections menu) and
Apply Corrections is set to Yes.
Key Path: AMPLITUDE / Y Scale, Y Axis Units
Remote Command:
See“Y Axis Units" on page 56.
Example:
UNIT:POW DBPT
dBG
Sets the amplitude units to dBG. This is a unit specifically applicable to EMI field strength
measurements. In the absence of a correction factor this unit is treated by the instrument
exactly as though it were dBmV. You must load an appropriate correction factor using