Technical data
574 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
27 Using the STATus System
Questionable Integrity Signal Positive Transition
:STATus:QUEStionable:INTegrity:SIGNal:PTRansition <number>
:STATus:QUEStionable:INTegrity:SIGNal:PTRansition?
Determines which bits in the Questionable Integrity Signal Condition register will set the
corresponding bit in the Questionable Integrity Signal Event register when the condition
register bit has a positive transition (0 to 1). The variable <number> is the sum of the
decimal values of the bits that you want to enable.
Key Type: There is no equivalent front-panel key.
Factory Preset: 32767 (all 1’s)
Range: 0 to 32767
Questionable Integrity Uncalibrated Register
Questionable Integrity Uncalibrated Condition
:STATus:QUEStionable:INTegrity:UNCalibrated:CONDition?
Returns the decimal value of the sum of the bits in the Questionable Integrity Uncalibrated
Condition register.
The data in this register is continuously updated and reflects the current conditions.
Key Type: There is no equivalent front-panel key.
Questionable Integrity Uncalibrated Enable
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle?
Determines which bits in the Questionable Integrity Uncalibrated Condition Register will set
bits in the Questionable Integrity Uncalibrated Event register, which also sets the Data
Uncalibrated Summary bit (bit 3) in the Questionable Integrity Register. The variable
<number> is the sum of the decimal values of the bits you want to enable.
Key Type: There is no equivalent front-panel key.
Factory Preset: 32767 (all 1’s)
Range: 0 to 32767
Questionable Integrity Uncalibrated Event Query
:STATus:QUEStionable:INTegrity:UNCalibrated[:EVENt]?