Technical data

Using the STATus System 27
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 573
The data in this register is continuously updated and reflects the current conditions.
Key Type: There is no equivalent front-panel key.
Questionable Integrity Signal Enable
:STATus:QUEStionable:INTegrity:SIGNal:ENABle <number>
:STATus:QUEStionable:INTegrity:SIGNal:ENABle?
Determines which bits in the Questionable Integrity Signal Condition Register will set bits in
the Questionable Integrity Signal Event register, which also sets the Integrity Summary bit
(bit 9) in the Questionable Register. The variable <number> is the sum of the decimal values
of the bits you want to enable.
Key Type: There is no equivalent front-panel key.
Factory Preset: 32767 (all 1’s)
Range: 0 to 32767
Questionable Integrity Signal Event Query
:STATus:QUEStionable:INTegrity:SIGNal[:EVENt]?
Returns the decimal value of the sum of the bits in the Questionable Integrity Signal Event
register.
The register requires that the associated PTR or NTR filters be set before a condition
register bit can set a bit in the event register.
The data in this register is latched until it is queried. Once queried, the register is cleared.
Key Type: There is no equivalent front-panel key.
Questionable Integrity Signal Negative Transition
:STATus:QUEStionable:INTegrity:SIGNal:NTRansition <number>
:STATus:QUEStionable:INTegrity:SIGNal:NTRansition?
Determines which bits in the Questionable Integrity Signal Condition register will set the
corresponding bit in the Questionable Integrity Signal Event register when the condition
register bit has a negative transition (1 to 0). The variable <number> is the sum of the
decimal values of the bits that you want to enable.
Key Type: There is no equivalent front-panel key.
Factory Preset: 0
Range: 0 to 32767