The register requires that the associated PTR or NTR filters be set before a condition
register bit can set a bit in the event register.
The data in this register is latched until it is queried. Once queried, the register is cleared.
Key Type: There is no equivalent front-panel key.
Questionable Integrity Negative Transition
:STATus:QUEStionable:INTegrity:NTRansition <number>
:STATus:QUEStionable:INTegrity:NTRansition?
Determines which bits in the Questionable Integrity Condition register will set the
corresponding bit in the Questionable Integrity Event register when the condition register
bit has a negative transition (1 to 0). The variable <number> is the sum of the decimal
values of the bits that you want to enable.
Key Type: There is no equivalent front-panel key.
Factory Preset: 0
Range: 0 to 32767
Questionable Integrity Positive Transition
:STATus:QUEStionable:INTegrity:PTRansition <number>
:STATus:QUEStionable:INTegrity:PTRansition?
Determines which bits in the Questionable Integrity Condition register will set the
corresponding bit in the Questionable Integrity Event register when the condition register
bit has a positive transition (0 to 1). The variable <number> is the sum of the decimal values
of the bits that you want to enable.
Key Type: There is no equivalent front-panel key.
Factory Preset: 32767 (all 1’s)
Range: 0 to 32767
Questionable Integrity Signal Register
Questionable Integrity Signal Condition
:STATus:QUEStionable:INTegrity:SIGNal:CONDition?
Returns the decimal value of the sum of the bits in the Questionable Integrity Signal
Condition register.