Technical data
568 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
27 Using the STATus System
Questionable Calibration Extended Failure Register
Questionable Calibration Extended Failure Condition
:STATus:QUEStionable:CALibration:EXTended:FAILure:
CONDition?
Returns the decimal value of the sum of the bits in the Questionable Calibration Extended
Failure Condition register.
The data in this register is continuously updated and reflects the current conditions.
Key Type: There is no equivalent front-panel key.
Questionable Calibration Extended Failure Enable
:STATus:QUEStionable:CALibration:EXTended:FAILure:ENABle
<number>
:STATus:QUEStionable:CALibration:EXTended:FAILure:ENABle?
Determines which bits in the Questionable Calibration Extended Failure Enable Register will
set bits in the Questionable Calibration Event register, which also sets the Calibration
Summary bit (bit 8) in the Questionable Register. The variable <number> is the sum of the
decimal values of the bits you want to enable.
Key Type: There is no equivalent front-panel key.
Example STAT:QUES:CAL:EXT:FAIL:ENABLE 16384 could be used if you
have turned off the automatic alignment and you want to query whether an alignment is
needed.
Factory Preset: 32767 (all 1’s)
Range: 0 to 32767
Questionable Calibration Extended Failure Event Query
:STATus:QUEStionable:CALibration:EXTended:FAILure[:EVENt]?
Returns the decimal value of the sum of the bits in the Questionable Calibration Extended
Failure Event register.
The register requires that the associated PTR or NTR filters be set before a condition
register bit can set a bit in the event register.
The data in this register is latched until it is queried. Once queried, the register is cleared.
Key Type: There is no equivalent front-panel key.